Routine maintenance of microplate reader
matters needing attention
※ The instrument shall be placed in an environment without magnetic field and interference voltage, which is lower than 40dB;
※ Avoid direct sunlight to prevent equipment aging;
※ During operation, the ambient temperature shall be between 15 ℃ and 40 ℃, and the ambient humidity shall be between 15% and 85%;
※ The operating voltage shall be stable during operation;
※ The air in the operating environment is clean to avoid water vapor and smoke;
※ Keep a dry, clean and horizontal working surface and a large enough operating space;
※ Use the liquid feeder to add liquid, and the liquid filler head cannot be mixed;
※ Try to use a washing machine to wash the board, so that it is clean and can effectively avoid cross contamination;
※ Operate in strict accordance with the instructions of the kit, and the reaction time is accurate;
※ During the measurement, please do not touch the enzyme label plate to prevent the operator's hand from being crushed during the transfer of the enzyme label plate;
※ Do not spill samples or reagents on the surface or inside of the instrument, and wash your hands after operation;
※ If the sample or reagent used has pollution, toxicity and biological hazards, please strictly follow the operating instructions of the kit to prevent damage to the operator;
※ If the instrument has been exposed to polluting or infectious substances, please wash and disinfect it;
※ Do not turn off the power during measurement;
※ For the deviation of the measurement results caused by the problems of the kit, the parameters should be modified in time according to the actual situation to achieve the best effect;
※ Use the rear cover to install the dust cover;
※ In case of technical failure, contact the manufacturer in time, and do not dismantle the microplate reader without authorization.